首页 | 本学科首页   官方微博 | 高级检索  
检索        


An interface to display special physiologic signals on patient monitors
Authors:Dr D John Doyle MD  PhD
Institution:(1) From the Department of Anaesthesia, The Toronto Hospital, Toronto General Division, Toronto, Ontario, Canada
Abstract:Objective. The objective of this study was to develop an interface to allow special physiologic signals (e.g., in a research setting) to be displayed on the invasive pressure channel of conventional clinical monitors. The interface accepts single-ended high-level signals for display using the pressure channel of patient monitors, which use strain-gauge transducers employing direct current (DC) excitation.Methods. By studying the electronic circuitry common to most clinical invasive pressure measurement systems (Wheatstone bridge, differential input instrumentation amplifier) it was possible to develop an interface to convert high-level single-ended signals into the low-level differential signal needed for input to an invasive pressure channel.Results and Conclusions. The device is useful when it is desired to display signals from special transducers on regular patient monitors. Schematic diagrams and sample results are provided.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏