Comparison of changes in retentive force of three stud attachments for implant overdentures |
| |
Authors: | Su-Min Kim Jae-Won Choi Young-Chan Jeon Chang-Mo Jeong Mi-Jung Yun So-Hyoun Lee Jung-Bo Huh |
| |
Affiliation: | Department of Prosthodontics, School of Dentistry, Pusan National University, Yangsan, Republic of Korea. |
| |
Abstract: | PURPOSEThe aim of this study was to compare the changes in retentive force of stud attachments for implant overdentures by in vitro 2-year-wear simulation.RESULTSA comparison of the initial retentive force revealed the highest value for Kerator, followed by the O-ring and EZ lock attachments. However, no significant difference was detected between Kerator and O-ring (P>.05). After 2500 insertion and removal cycles, the highest retention loss was recorded for O-ring, and no significant difference between Kerator and EZ lock (P>.05). Also, Kerator showed the highest retentive force, followed by EZ lock and O-ring, after 2500 cycles (P<.05). Based on SEM analysis, the polymeric components in O-ring and Kerator were observed to exhibit surface wear and deformation.CONCLUSIONAfter 2500 insertion and removal cycles, all attachments exhibited significant loss in retention. Mechanism of retention loss can only be partially explained by surface changes. |
| |
Keywords: | Overdenture attachment Resilient attachment Surface wear Retention |
|
|