首页 | 本学科首页   官方微博 | 高级检索  
     


Clinical use of stereoscopic X-ray positioning of patients treated with conformal radiotherapy for prostate cancer
Authors:Soete Guy  Verellen Dirk  Michielsen Dirk  Vinh-Hung Vincent  Van de Steene Jan  Van den Berge Dirk  De Roover Patricia  Keuppens Frans  Storme Guy
Affiliation:Department of Radiotherapy, Academic Hospital Free University of Brussels, Brussels, Belgium. guy.soete@az.vub.ac.be
Abstract:
PURPOSE: To evaluate accuracy and time requirements of a stereoscopic X-ray-based positioning system in patients receiving conformal radiotherapy to the prostate. METHODS AND MATERIALS: Setup errors of the isocenter with regard to the bony pelvis were measured by means of orthogonal verification films and compared to conventional positioning (using skin drawings and lasers) and infrared marker (IR) based positioning in each of 261 treatments. In each direction, the random error represents the standard deviation and the systematic error the absolute value of the mean position. Time measurements were done in 75 treatments. RESULTS: Random errors with the X-ray positioning system in the anteroposterior (AP), lateral, and longitudinal direction were (average +/- 1 standard deviation) 2 +/- 0.6 mm, 1.7 +/- 0.6 mm, and 2.4 +/- 0.7 mm. The corresponding values of conventional as well as IR positioning were significantly higher (p < 0.01). Systematic errors for X-ray positioning were 1.1 +/- 1.2 mm AP, 0.6 +/- 0.5 mm laterally, and 1.5 +/- 1.6 mm longitudinally. Conventional and IR marker-based positioning showed significantly larger systematic errors AP and laterally, but longitudinally, the difference was not significant. Depending on the axis looked at, errors of >or=5 mm occurred in 2%-14% of treatments after X-ray positioning, 13%-29% using IR markers, and 28%-53% with conventional positioning. Total linac time for one treatment session was 14 min 51 s +/- 4 min 18 s, half of which was used for the X-ray-assisted positioning procedure. CONCLUSION: X-ray-assisted patient positioning significantly improves setup accuracy, at the cost of an increased treatment time.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号