首页 | 本学科首页   官方微博 | 高级检索  
     


Retrieval analysis of porous-coated components for total knee arthroplasty. A report of two cases
Authors:C S Ranawat  N A Johanson  C M Rimnac  T M Wright  R E Schwartz
Abstract:
Retrieval analysis of two total knee replacements incorporating porous-coated components for biologic fixation was performed. The significant finding in a 63-year-old woman was fracture of the porous-coated metal backing of the tibial component, secondary to aseptic loosening. The significant findings at the time of revision surgery in an 82-year-old woman were nonuniform bone ingrowth (primarily around fixation pegs) and fracture of metal beads from the metal backing of the patellar component.