首页 | 本学科首页   官方微博 | 高级检索  
     


Development of a lateral-flow immunochromatographic test device for the rapid detection of difloxacin residues
Authors:Ai-min Zhi  Bin-bin Li  Qing-tang Liu  Xiao-fei Hu  Dong Zhao  Yu-ze Hou
Affiliation:1. Key Laboratory of Animal Immunology , Henan Academy of Agricultural Sciences , Zhengzhou, 450002, P.R. China;2. Key Laboratory of Animal Immunology , Henan Academy of Agricultural Sciences , Zhengzhou, 450002, P.R. China;3. Jiaozuo Entry-Exit Inspection and Quarantive Bureau of P.R. China , Henan, 473036, P.R. China
Abstract:
In this study, a rapid competitive immunochromatographic test strip has been developed for specifically testing residues of difloxacin (DIF). 1H10-2B2, one of the three hybridoma lines that were tested and selected by enzyme linked immunosorbent assay, was used in the test strip. The monoclonal antibody has a good sensitivity with an IC50 of 2.2 ng/mL to DIF, 0.24% cross-reactivity towards danofloxacin and no cross-reactivity towards other related compounds and other drugs. The calibration curve of DIF test strip presented a typical sigmoidal curve. In practice, the lower limit of detection using a strip reader was 0.5 ng/ml while 2 ng/ml by unaided visual assessment. Samples of milk were spiked at 2–8 ng/mL, presenting recoveries between 94.5 and 107%, and the coefficient of variation (CV,%) (1.6–9%). The data above suggests that the strip meets the requirements of high sensitivity, good specificity, simplicity and speed, as well as the characteristics of reproducibility and accuracy.
Keywords:difloxacin  test devices  colloidal gold  immunochromatography  rapid test  immunoassays
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号