Automatic adjustment of pacing output in the clinical setting |
| |
Authors: | Ribeiro Antonio Luiz P Rincón Leonor G Oliveira Bruna G Vinha Caio R Melatto Douglas Torres Ana A Barros Vladimir C V Levine Paul A |
| |
Affiliation: | a Pacemaker Laboratory, Hospital das Clínicas, Belo Horizonte, Brazil b Internal Medicine Department, School of Medicine, Universidade Federal de Minas Gerais, Belo Horizonte, Brazil c St. Jude Medical Brasil, São Paulo, Brazil d St. Jude Medical CRMD, Sylmar, Calif, USA |
| |
Abstract: | ![]()
BackgroundAutoCapture (AC) is a programmable feature that enables the pacemaker to both track the capture threshold and automatically adjust the output on a beat-by-beat basis. Although AC safely and significantly reduces the current drainage, some authors have argued that the longevity benefit of such a system is overstated. This study aims to estimate the longevity extension that can be obtained, in the clinical routine, by turning the AC on in comparison to pacemakers programmed to operate at the shipped and manually optimized output.MethodsWe selected 83 consecutive patients who received implanted St Jude's Affinity pacemakers >6 months earlier. Eight patients died or were lost to follow-up and in 9 subjects the AC could not be turned on. In the remaining 66 patients, current drain and estimated longevity were compared in 3 situations: (1) AC on; (2) AC off, optimized programming (100%-150% voltage threshold); (3) AC off, shipped output (3.5 V).ResultsFive patients had large variations (>1 V) of the AC threshold. Current drainage was 8.0 ± 0.9 mA in the AC group, 8.7 ± 1.8 mA with AC off and optimized programming, and 11.3 ± 2.3 mA at shipped output (P < .01). Estimated longevity was significantly extended (P < .01) by AC (12.1 ± 1.0 years) when compared to shipped (8.9 ± 1.7 years) and optimized programming (11.3 ± 1.4 years).ConclusionReprogramming the pacemaker output significantly enhanced its estimated longevity; AC added a moderate but significant extension over manual reprogramming and was associated with increased safety in patients with large ventricular threshold variations. |
| |
Keywords: | |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|