2015年广西数字化摄影受检者入射体表剂量水平调查与分析
引用本文:梁挺,陈掌凡,陈发想,董颖,吴应宇,梁桂强,赵新春,谢萍,覃志英. 2015年广西数字化摄影受检者入射体表剂量水平调查与分析[J]. 中国辐射卫生, 2016, 25(5): 516-519. DOI: 10.13491/j.cnki.issn.1004-714x.2016.05.002
作者姓名:梁挺  陈掌凡  陈发想  董颖  吴应宇  梁桂强  赵新春  谢萍  覃志英
作者单位:广西壮族自治区疾病预防控制中心, 放射卫生防护所, 广西南宁 530028
基金项目:广西自然科学基金(2014GXNSFAA118223,2014GXNSFBA118173)
摘    要:
目的 了解广西地区数字化摄影(DR)所致受检者的入射体表剂量水平。方法 采用分层随机抽样方法,选取广西区内14家医院不同时间段进行DR检查的受检者,使用热释光剂量计测量受检者不同部位的入射体表剂量。结果 广西地区一级医院的头颅(PA)、头颅(LAT)、胸部(PA)、胸部(LAT)、胸椎(AP)、腰椎(AP)6个部位入射体表剂量的75%位值超出了国家医疗照射指导水平;二级和三级医院的胸部(PA)和胸部(LAT)两个部位入射体表剂量的75%位值超出国家医疗照射指导水平。结论 广西地区DR摄影的入射体表剂量较高,应进一步加强受检者的入射体表剂量监测。

关 键 词:数字化摄影  入射体表剂量  医疗照射指导水平  
收稿时间:2016-07-10

Investigation and Analysis About Entrance Surface Dose Induced by Medical Digital Radiography in Guangxi 2015
LIANG Ting,CHEN Zhang-fan,CHEN Fa-xiang,DONG Yin,WU Ying-yu,LIANG Gui-qiang,ZHAO Xin-chun,XIE Ping,QIN ZHi-ying. Investigation and Analysis About Entrance Surface Dose Induced by Medical Digital Radiography in Guangxi 2015[J]. Chinese Journal of Radiological Health, 2016, 25(5): 516-519. DOI: 10.13491/j.cnki.issn.1004-714x.2016.05.002
Authors:LIANG Ting  CHEN Zhang-fan  CHEN Fa-xiang  DONG Yin  WU Ying-yu  LIANG Gui-qiang  ZHAO Xin-chun  XIE Ping  QIN ZHi-ying
Affiliation:Guang xi Zhuang Autonomous Region Center for Disease Control and Prevention, Nanning 530028 China
Abstract:
Objective To acquire information about the entrance surface dose level induced by medical digital radiography (DR) in Guangxi. Methods Stratified random sampling method is used. We choose the subjects examined by DR on different time in fourteen hospitals in Guangxi, using thermoluminescence dosimeter to measure subjects' entrance surface dose in different parts. Results There are head (PA)、head(LAT)、chest (PA)、chest (LAT)、thoracicvertebra (AP)、lumbarvertebra (AP) of which the amount to six parts' entrance surface dose of 75% bit value exceeds the national guidance levels for medical exposure in the grade-one hospitals; there are chest (PA) andchest(LAT) amount to two parts' entrance surface dose of 75% bit value exceeds the national guidance levels for medical exposure in the grade-two hospitals and grade-three hospitals. Conclusion The entrance surface dose of DR photography in Guangxi is hight. We should a further strengthen the monitoring plan about the entrance surface dose.
Keywords:Medical Digital Radiography  Entrance Surface Dose  Guidance Levels for Medical Exposure  