A rapid, objective method for the detection of lymphocytotoxic antibodies using flow cytometry |
| |
Authors: | D Talbot B K Shenton A L Givan G Proud R M Taylor |
| |
Affiliation: | 1. Anatomy, The Medical School, University of Newcastle upon Tyne, U.K.;2. Surgery Departments, The Medical School, University of Newcastle upon Tyne, U.K. |
| |
Abstract: | Whilst several centres have reported lymphocytotoxic antibody detection using single and dual fluorescent stains with analysis of the fluorescence emitted from the cell population present in a well of a multiwell plate, problems are encountered with cell concentration and light emission overlap. A method we have developed using flow cytometry produces similar values of percentage cell death using single or double staining techniques (correlation coefficient = 0.9896). This method is not influenced by slight variation in cell number or light emission overlap. The effect of introducing red cell impurities into the normal lymphocyte preparation is described. |
| |