Authors:D Talbot  B K Shenton  A L Givan  G Proud  R M Taylor
Affiliation:1. Anatomy, The Medical School, University of Newcastle upon Tyne, U.K.;2. Surgery Departments, The Medical School, University of Newcastle upon Tyne, U.K.
Abstract:
Whilst several centres have reported lymphocytotoxic antibody detection using single and dual fluorescent stains with analysis of the fluorescence emitted from the cell population present in a well of a multiwell plate, problems are encountered with cell concentration and light emission overlap. A method we have developed using flow cytometry produces similar values of percentage cell death using single or double staining techniques (correlation coefficient = 0.9896). This method is not influenced by slight variation in cell number or light emission overlap. The effect of introducing red cell impurities into the normal lymphocyte preparation is described.