首页 | 本学科首页   官方微博 | 高级检索  
检索        


Inadvertent stenting of patent ductus arteriosus with Amplatzer Vascular Plug.
Authors:Alexander J Javois  Tarek S Husayni  David Thoele  Andrew H Van Bergen
Institution:The Heart Institute for Children, Advocate Hope Children's Hospital, Oak Lawn 60453, and Rush Medical College, Chicago, Illinois, USA. alexjavois@thic.com
Abstract:We report a 12-month old patient who presented for murmur evaluation after percutaneous closure of type C patent ductus arteriosus (PDA) using a 10 mm Amplatzer Vascular Plug (AGA Medical Corporation, Golden Valley, MN) at an outside institution. Echocardiography revealed a large left-to-right shunt through the implanted device, inadvertently stenting the PDA instead of closing it. The patient underwent repeat catheterization with successful coil implantation within the Amplatzer Vascular Plug, completely eliminating the large residual ductal shunt. Although challenging, this case illustrates the technique of implanting coils within this occlusion device. This case also illustrates that occlusion of type C PDA utilizing the Amplatzer Vascular Plug may not only result in incomplete occlusion but also create a potentially worse clinical situation in which the PDA is stretched larger and stented open. Without consideration of simultaneous coil implantation within this device, use of the Amplatzer Vascular Plug might be contraindicated in type C PDA, because there may be no way to ensure successful closure by the Vascular Plug alone.
Keywords:Amplatzer Vascular Plug  patent ductus arteriosus  PDA
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号