首页 | 本学科首页   官方微博 | 高级检索  
     


Biomechanical evaluation of the less invasive stabilization system and the 95-degree angled blade plate for the internal fixation of distal femur Fractures in human cadaveric bones with high bone mineral density
Authors:Zlowodzki Michael  Williamson Scott  Zardiackas Lyle D  Kregor Philip J
Affiliation:Department of Orthopedic Surgery, University of Minnesota, Minneapolis, USA.
Abstract:
BACKGROUND: The less invasive stabilization system (LISS) is an internal fixator that utilizes unicortical locked screws for fixation of distal femur fractures. A question is whether locked unicortical screw fixation is sufficient, when compared with a standard implant such as a blade plate. METHODS: Eight matched pairs of fresh-frozen cadaveric femora were instrumented with either the LISS or a 95-degree blade plate. A 4-cm supracondylar gap fracture model was created and all bone-implant constructs were tested to failure in axial loading. RESULTS: All constructs failed by plastic deformation of the implant. There was no significant difference between the LISS and the blade plate constructs with respect to load to failure. CONCLUSIONS: Despite unicortical fixation axial loading to failure of the LISS did not result in implant/screw pull-out neither proximally nor distally. However, there does not appear to be a biomechanical advantage of using the LISS as opposed to a blade plate in bones with high bone mineral density.
Keywords:
本文献已被 PubMed 等数据库收录!