首页 | 本学科首页   官方微博 | 高级检索  
     


Clinical use of the abiomed BVS 5000 as a pulsatile extracorporeal membrane oxygenation unit
Authors:Samuels Louis  Entwistle John  Holmes Elena  Eaton Doug  Thomas Matthew P  Menzano Gina  Promisloff Robert
Affiliation:The Lankenau Hospital, Department of Cardiothoracic Surgery, Wynnewood, PA, USA.
Abstract:
The traditional extracorporeal membrane oxygenation circuit uses a centrifugal pump. These pumps require close monitoring and are subject to complications. In addition, they do not take advantage of the potential benefits of pulsatile flow. These extracorporeal membrane oxygenation circuits use a single pump with an inline oxygenator. If cardiac failure persists after respiratory recovery has occurred, removal of the oxygenator requires an additional procedure to convert the patient to biventricular support. This report describes a circuit in which an oxygenator is connected to a pulsatile ventricular assist device. Single and dual circuit configurations are illustrated. Recommendations for pulmonary care during support are also described.
Keywords:
本文献已被 PubMed 等数据库收录!