Evaluation of the URIT-2900 automated hematology analyzer for screening of thalassemia and hemoglobinopathies in Southeast Asian populations
Authors:Karnpean Rossarin  Pansuwan Anupong  Fucharoen Goonnapa  Fucharoen Supan
Institution:The Graduate School, Faculty of Associated Medical Sciences, Khon Kaen University, Khon Kaen, Thailand.
Abstract:BackgroundThe effectiveness of the URIT-2900 Hematology Analyzer for screening of hemoglobinopathies commonly found in Southeast Asian populations was examined.MethodsAppropriate cut-off values of MCV and MCH for screening of α0 and β thalassemias were derived from the receiver operator characteristic curve conducted initially on 279 subjects with various thalassemia genotypes. Validation was performed additionally in a cohort of another unrelated 313 subjects.ResultsThe best cut off values of MCV and MCH were found to be 78 fL and 27 pg, respectively. Using these cut off values in combination with the dichlorophenolindophenol test in screening of α0 thalassemia, β thalassemia and Hb E in a cohort study revealed 100% sensitivity, 79.6% specificity, 80.0% positive predictive value and 100% negative predictive value.ConclusionThe combined blood cell counting using the URIT-2900 Automated Hematology Analyzer and dichlorophenolindophenol test is suitable for population screening of thalassemia and hemoglobinopathies in Southeast Asia.