首页 | 本学科首页   官方微博 | 高级检索  
     


Assaying 192Ir line sources using a standard length well chamber
Authors:Brezovich Ivan A  Popple Richard A  Duan Jun  Shen Sui  Pareek Prem N
Affiliation:Department of Radiation Oncology, University of Alabama at Birmingham, Birmingham, Alabama 35233, USA. ibrezovich@uabmc.edu
Abstract:
The strength of intravascular 192Ir sources is typically measured by the manufacturer before shipment, and treatment planning is based on that assay. However, in-house verification of source strength is required at some institutions by state law or internal policy, is recommended by the AAPM TG 60 report on intravascular brachytherapy, and is considered a necessity by many medical physicists. To accommodate the long sources used in intravascular therapy, special well chambers with extended regions of constant response have been designed. To allow assays using a widely available standard well chamber, we have measured its position dependent sensitivity and derived from it a table of correction factors that account for the extended length of intravascular sources. An experimental verification shows that application of these correction factors yields assays with sufficient accuracy for routine quality assurance tests.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏