Long-term biocompatibility of implanted polymer-based intrafascicular electrodes |
| |
Authors: | Lawrence Stephen M Larsen Jytte O Horch Kenneth W Riso Ron Sinkjaer Thomas |
| |
Affiliation: | Department of Bioengineering, University of Utah, Salt Lake City, Utah 84112, USA. |
| |
Abstract: | Polymer-based longitudinal intrafascicular electrodes (polyLIFEs) were chronically implanted into the sciatic nerve of white New Zealand rabbits (n=8) for a period of 6 months (hereafter referred to as the long-term group). The impact of the implantation procedure, as observed 6 months post surgery, was evaluated in a sham-treated control group (n=9). The contralateral sciatic nerve served as the control for each animal. Nerve-fiber counts, fiber diameters, and myelin thickness were estimated at the level of the implant site, 1.5 cm proximally, and 1.5 cm distally for both nerves in sham-treated and long-term groups. Implantation of polyLIFEs had no significant effect on fiber counts, nerve-fiber diameter, or myelin thickness. A slight increase in connective tissue in the vicinity of the implant site was evident in the long-term group, including a thin but dense capsule immediately surrounding the implanted electrode. |
| |
Keywords: | |