首页 | 本学科首页   官方微博 | 高级检索  
检索        

OBI图像引导在鼻咽癌IMRT中的应用
引用本文:钟嘉健,王成涛,邓永锦,丘敏敏,林泽煌,闵蒙真,肖振华,王振宇,文碧秀.OBI图像引导在鼻咽癌IMRT中的应用[J].中华放射肿瘤学杂志,2017,26(9):992-996.
作者姓名:钟嘉健  王成涛  邓永锦  丘敏敏  林泽煌  闵蒙真  肖振华  王振宇  文碧秀
作者单位:510080 广州,中山大学附属第一医院放疗科
摘    要:目的 探索IGRT在保障鼻咽癌IMRT精确性中的应用。方法 应用直线加速器OBI系统对10例鼻咽癌患者实施IMRT,观察并记录每次摆位后、体位调整后、治疗后的IGRT,分析其影像与CT模拟定位影像之间的摆位差异。结果 摆位后左右、头脚、前后方向误差分别为(0.22±1.00)、(-0.37±1.28)、(0.04±1.36) mm;体位调整后对应为(0.29±0.76)、(-0.04±0.78)、(-0.01±0.92) mm;治疗后对应为(0.20±0.78)、(0.16±0.80)、(0.05±0.92) mm。摆位后左右、头脚、前后方向误差≤1 mm概率分别为81.0%、77.6%、88.2%;体位调整后对应为92.5%、96.4%、96.4%;治疗后对应为91.7%、94.9%、96.8%。结论 鼻咽癌患者实施分次IMRT过程中进行OBI系统下的精准IGRT技术非常重要,必要时需参照摆位后的IGRT影像调整患者的放疗体位,修正摆位误差,有效提高分次IMRT的精确性。

关 键 词:鼻咽肿瘤/调强放射疗法  摆位误差  影像学  
收稿时间:2017-02-07

Application of OBI system in intensity-modulated radiotherapy for nasopharyngeal carcinoma
Zhong Jiajian,Wang Chentao,Deng Yongjin,Qiu Minmin,Lin Zehuang,Min Mengzhen,Xiao Zhenhua,Wang Zhenyu,Wen Bixiu.Application of OBI system in intensity-modulated radiotherapy for nasopharyngeal carcinoma[J].Chinese Journal of Radiation Oncology,2017,26(9):992-996.
Authors:Zhong Jiajian  Wang Chentao  Deng Yongjin  Qiu Minmin  Lin Zehuang  Min Mengzhen  Xiao Zhenhua  Wang Zhenyu  Wen Bixiu
Institution:Department of Radiation Oncology,First Affiliated Hospital of Sun Yat-sen University,Guangzhou 510080China
Abstract:Objective To examine the application of On-Board Imaging (OBI) system-based image-guided radiotherapy (IGRT) in the improvement of the precision of intensity-modulated radiotherapy (IMRT) for nasopharyngeal carcinoma.Methods Ten patients with nasopharyngeal carcinoma were treated with IMRT using the OBI system. The IGRT images after positioning, position adjustment, and treatment were observed and recorded to investigate the image difference between CT simulation and IGRT. Results The deviations in the x (lateral), y (cranial-caudal), and z (ventral-dorsal) directions between CT simulation and IGRT images were 0.22±1.00 mm,-0.37±1.28 mm, and 0.04±1.36 mm, respectively, after positioning, 0.29±0.76 mm,-0.04±0.78 mm, and -0.01±0.92 mm, respectively, after position adjustment, and 0.20±0.78 mm, 0.16±0.80 mm, and 0.05±0.92 mm, respectively, after treatment. The probabilities of a ≤1 mm deviation in the x, y, and z directions were 81.0%, 77.6%, and 88.2%, respectively, after positioning, 92.5%, 96.4%, and 96.4%, respectively, after position adjustment, and 91.7%, 94.9%, and 96.8%, respectively, after treatment. Conclusions The application of OBI system-based IGRT is very important in the improvement of the precision of fractionated IMRT for patients with nasopharyngeal carcinoma. The position of the patient should be adjusted based on the IGRT image after positioning in order to correct set-up error and effectively increase the precision of fractionated IMRT.
Keywords:Nasopharyngeal neoplasms/intensity-modulated radiotherapy  Set-up error  Iconography
本文献已被 万方数据 等数据库收录!
点击此处可从《中华放射肿瘤学杂志》浏览原始摘要信息
点击此处可从《中华放射肿瘤学杂志》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号