Authors:Kyungnam Park MD  Jangyoung Lee MD PhD  Soo‐young Kim MD PhD  Jinwoo Kim MS  Insoo Kim MS  Seung Pill Choi MD PhD  Sikyung Jeong MD PhD  Sungyoup Hong MD PhD
Affiliation:1. Department of Emergency Medicine, Eulji University, , Daejeon, Korea;2. Occupational and Environmental Medicine, Eulji University, , Daejeon, Korea;3. Department of Emergency Medical Service Technology, Daejeon Health Science College, , Daejeon, Korea;4. Department of Emergency Medicine, Daejeon St. Mary's Hospital, The Catholic University of Korea, , Daejeon, Korea;5. Department of Emergency Medicine, Yeoeudo St. Mary's Hospital, The Catholic University of Korea, , Seoul, Korea
Abstract:
This study assessed the method of fluid infusion control using an IntraVenous Infusion Controller (IVIC). Four methods of infusion control (dial flow controller, IV set without correction, IV set with correction and IVIC correction) were used to measure the volume of each technique at two infusion rates. The infused fluid volume with a dial flow controller was significantly larger than other methods. The infused fluid volume was significantly smaller with an IV set without correction over time. Regarding the concordance correlation coefficient (CCC) of infused fluid volume in relation to a target volume, IVIC correction was shown to have the highest level of agreement. The flow rate measured in check mode showed a good agreement with the volume of collected fluid after passing through the IV system. Thus, an IVIC could assist in providing an accurate infusion control.
Keywords:equipment and supplies  intravenous infusion  medication error  medication management  medication therapy
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号