首页 | 本学科首页   官方微博 | 高级检索  
     


Concentric needle electrode for neuromuscular jitter analysis
Authors:Ertaş M  Baslo M B  Yildiz N  Yazici J  Oge A E
Affiliation:Department of Neurology and Electrodiagnostic Neurology, Istanbul University, Faculty of Medicine, Capa, Istanbul 34390, Turkey. mertas@turk.net
Abstract:We used a concentric needle electrode (CNE) with 2 kHZ low-cut filter and a single fiber electrode (SFE) in the same subjects for neuromuscular jitter measurement in the extensor digitorum communis (EDC) and orbicularis oculi (OOc) muscles. At the same session, 20 jitter values were obtained from each subject with each electrode. For EDC (during voluntary contraction), mean jitter values with SFE and CNE were 23.4 +/- 8 micros and 23.3 +/- 8 micros in 10 normals; and 56.8 +/- 28 micros and 57.4 +/- 33 micros in 10 myasthenics. For OOc (during electrical stimulation), mean jitter values with SFE and CNE were 17.9 +/- 5 micros and 16.3 +/- 4 micros in 11 normal subjects, and 41.2 +/- 29 micros and 36.7 +/- 27 micros in 10 myasthenics. For both muscles, the numbers of individual abnormal jitter values with SFE and CNE were highly comparable. Both needles labeled the same patients as having "normal" or "abnormal" neuromuscular transmission. CNE may be an alternative to SFE in neuromuscular jitter analysis.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号