Needle position during (125)I seed implantation: accurately recognized by sagittal transrectal ultrasonography [corrected |
| |
Authors: | Ishiyama Hiromichi Kotani Shouko Satoh Takefumi Uemae Mineko Baba Shiro Hayakawa Kazushige |
| |
Affiliation: | (1) Department of Radiology, Kitasato University School of Medicine, 1-15-1 Kitasato, Sagamihara 228-8555, Japan;(2) Department of Urology, Kitasato University School of Medicine, Sagamihara, Japan;(3) Division of Radiation Oncology, Kitasato University Hospital, Sagamihara, Japan |
| |
Abstract: | Purpose The aim of this study was to assess the variation of probe rotation angles for detecting a single needle using sagittal images of transrectal ultrasonography (TRUS). Materials and methods A phantom study was performed. One needle was inserted through each of 10 holes of the template, and variations in the probe rotation angles for detecting the needle were measured. Results The mean variation of probe rotation for detecting a single needle was 17.0° (range 4°–25°). Slightly broader variation was seen for the needle in holes farther away from the probe. Conclusion Probe rotation angles for detecting a single needle displayed considerable variation. Seed locations recognized on sagittal imaging by TRUS are thus indeterminate, and real-time dose calculations using TRUS for 125I seed implantation should be used with care. |
| |
Keywords: | Prostate cancer Brachytherapy 125I Transrectal ultrasonography Real-time dose calculation |