摘    要:
我院在1974~1992年期间共植入永久性心脏起搏器400例,其中5例发生了心内膜电极导管损伤、断裂。其原因主要是医源性损伤4例和电极老化1例。为了预防,提出训练有素的手术者,选用高质量的起搏系统、避免患肢过强活动是减少心内膜电极导管损伤断裂的主要措施。

关 键 词:心内膜电极,损伤,断裂

ANALYSIS OF ENDOCARDIAL ELECTRODE FRACTURE IN 5 CASES
Zhang Quan-Fa,Cui Chang-Zong,Fu Wen,et al.. ANALYSIS OF ENDOCARDIAL ELECTRODE FRACTURE IN 5 CASES[J]. Chinese Circulation Journal, 1994, 0(4)
Authors:Zhang Quan-Fa  Cui Chang-Zong  Fu Wen  et al.
Abstract:
Four hundred cases were implanted artificial pacemaker during 1974 to 1992 in our hospital,5 cases of electrode fracture occurred,caused by iatrogenic lead damage in 4 cases,by ageing of electrode in 1.In order to prevent this complication,it is suggested that well trained operator,careful manipulation,good quality of the pacing system,and restriction of limb movementshould be emphasized.
Keywords:Endocardial electrode  Damagr  Fracture  