Abstract:
A perfusion system was constructed which allows the fast application of different solutes underneath a water immersion objective. The perfusion system is mounted into the immersion objective by milling a slot into the frontal metal plate of the lens holder. It consists of a five-channel pipette fixed to the objective and solution reservoirs gated by computer controlled magnetic valves. Up to five different solutions can be applied to the specimen under study. The solution between objective and specimen is completely exchanged after 1–2 s as determined from fluorescence measurements. This arrangement is optimized for [Ca2+] measurements with a fluorescence measurement system in tissue slices, where upright microscopes are required. It offers the advantage of saving a micromanipulator for the perfusion pipette and facilitates a fast, reproducible and precise positioning of the perfusion system.
Keywords:Fast perfusion system  Multi-channel pipette  Calcium  Fura-2 fluorescence measurements  Upright microscope  Slice  Substance P