Evaluating the use and acceptability of a needle-remover device in India
Authors:Muller Nancy  Steele Matthew  Balaji K A  Krishna Madhu  Berman Adriane P  Robertson Joan  Vail Janet
Affiliation:PATH, Seattle, WA, USA. nmuller@path.org
Abstract:
The objective of this study was to assess the effect of the use of a manually operated needle remover on sharps-waste management practices in clinical settings in India - specifically, evaluating its acceptability and performance. Thirty-one Balcan Mini-Destructor needle removers were introduced into seven health facilities in two cities in India - Delhi and Jaipur. One hundred and nineteen health workers, including auxiliary nurse midwives, nurses, and laboratory staff, used the device. Data were prospectively collected by observation and interview on device usage, malfunction and acceptability over a 23-week period. Focus group discussions on current practices were conducted prior to study initiation and, after completion, on device acceptability and performance. The manual needle remover was well accepted. Devices were seen as easy to use and durable. In total, 88,719 needles were removed. In conclusion, the needle-remover device was considered an acceptable method of preventing needle reuse and isolating infectious sharps waste in clinical settings.