Use of the bougie in simulated difficult intubation. 2. Comparison of single-use bougie with multiple-use bougie |
| |
Authors: | Marfin A G Pandit J J Hames K C Popat M T Yentis S M |
| |
Affiliation: | Nuffield Department of Anaesthesia, The John Radcliffe Hospital, Oxford, OX3 9 DU. |
| |
Abstract: | We studied the success rates for tracheal intubation in 32 healthy, anaesthetised patients during simulated grade IIIa laryngoscopy, randomised to either the multiple-use or the single-use bougie. Success rates (primary end-point) and times taken (secondary end-point) to achieve tracheal intubation were recorded. The multiple-use bougie was more successful than the single-use one (15/16 successful intubations vs. 9/16; p = 0.03). With either device, median [range] total tracheal intubation times for successful attempts were < 54 [24-84] s and there were no clinically important differences between these times. We conclude that the multiple-use bougie is a more reliable aid to tracheal intubation than the single-use introducer in grade IIIa laryngoscopy. |
| |
Keywords: | Intubation tracheal difficult Equipment multiple‐use and single‐use tracheal tube introducers |
本文献已被 PubMed 等数据库收录! |