PET performance measurements for an LSO-based combined PET/CT scanner using the National Electrical Manufacturers Association NU 2-2001 standard.
Authors:Yusuf E Erdi  Sadek A Nehmeh  Tim Mulnix  John L Humm  Charles C Watson
Affiliation:Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, New York 10021, USA. erdiy@mskcc.org
Abstract:
Results of performance measurements for a lutetium oxyorthosilicate (LSO)-based PET/CT scanner using new National Electrical Manufacturers Association (NEMA) NU 2-2001 standards are reported. METHODS: Performance measurements following the NU 2-2001 standards were performed on an LSO-based PET/CT scanner. In addition, issues associated with the application of the NEMA standard to LSO-based tomographs in the presence of intrinsic radiation are discussed. RESULTS: We report on some difficulties experienced in following the suggested NEMA measurement techniques and describe alternative approaches. Measurements with the new standard (as compared with NU-1994) incorporate the effects of activity outside the scanner and facilitate measurements of the entire axial field of view. Realistic clinical conditions are also simulated in image quality measurements of a torso phantom. CONCLUSION: We find that, with appropriate modifications, NU 2-2001 can be successfully applied to LSO-based scanners.