千伏级CBCT图像CT值校正及在放疗剂量计算中应用
引用本文:王雪桃,柏森,李光俊,蒋晓芹,苏晨,李衍龙,朱智慧. 千伏级CBCT图像CT值校正及在放疗剂量计算中应用[J]. 中华放射肿瘤学杂志, 2015, 24(4): 457-461. DOI: 10.3760/cma.j.issn.1004-4221.2015.04.026
作者姓名:王雪桃  柏森  李光俊  蒋晓芹  苏晨  李衍龙  朱智慧
作者单位:430072 武汉大学物理科学与技术学院(王雪桃、李衍龙、苏晨);610064成都,四川大学华西医院肿瘤放射物理技术中心(柏森、李光俊、蒋晓芹、朱智慧)
摘    要:目的 研究千伏级CBCT图像CT值校正方法,提高其用于剂量计算的准确性。方法 以扇形束计划CT作为先验信息,将CBCT与计划CT图像进行刚性配准,通过将CBCT与计划CT图像相减得到CBCT散射背景估计,对散射背景进行低通滤波处理,最后将原始CBCT图像减去滤波后的散射背景得到校正的CBCT图像。对Catphan600模体和4例盆腔恶性肿瘤患者的CBCT图像进行校正,配对t检验校正前后CBCT与计划CT的差异,评估校正后的CBCT图像质量并分析用于剂量计算的准确性。 结果 经CT值校正后CBCT图像伪影明显减少,空气、脂肪、肌肉、股骨头的平均值校正前与计划CT分别相差232、89、29、66 HU,而校正后平均值差别缩小至5 HU内(P=0.39、0.66、0.59、1.00)。校正后CBCT图像用于剂量计算误差在2%内。 结论 校正后的CBCT图像CT值与计划CT的CT值相似,用于剂量计算可得到准确的结果。

关 键 词:体层摄影术  X线计算机  锥形束  CT值校正  剂量计算  
收稿时间:2013-12-30

Investigation of CT numbers correction of kilo-voltage cone-beam CT images for accurate dose calculation
Wang Xuetao,Bai Sen,Li Guangjun,Jiang Xiaoqin,Su Chen,Li Yanlong,Zhu Zhihui. Investigation of CT numbers correction of kilo-voltage cone-beam CT images for accurate dose calculation[J]. Chinese Journal of Radiation Oncology, 2015, 24(4): 457-461. DOI: 10.3760/cma.j.issn.1004-4221.2015.04.026
Authors:Wang Xuetao  Bai Sen  Li Guangjun  Jiang Xiaoqin  Su Chen  Li Yanlong  Zhu Zhihui
Affiliation:School of Physics and Technology, Wuhan University, Wuhan 430072, China
Abstract:Objective To study CT numbers correction of kilo-voltage cone-beam CT (KV-CBCT) images for dose calculation. Method Aligning the CBCT images with plan CT images, then obtain the background scatter by subtracting CT images from CBCT images. The background scatter is then processed by low-pass filter. The final CBCT images are acquired by subtracting the background scatter from the raw CBCT. KV-CBCT images of Catphan600 phantom and four patients with pelvic tumors were obtained with the linac-integrated CBCT system. The CBCT images were modified to correct the CT numbers. Finally, compare HU numbers between corrected CBCT and planning CT by paired T test. Evaluate the image quality and accuracy of dose calculation of the modified CBCT images. Results The proposed method reduces the artifacts of CBCT images significantly. The differences of CT numbers were 232 HU, 89 HU, 29 HU and 66 HU for air, fat, muscle and femoral head between CT and CBCT respectively (P=0.39,0.66,0.59,1).The differences of CT numbers between CT and CBCT was reduced to within 5 HU. And the error of dose calculation with corrected CBCT images was within 2%. Conclusions The CT numbers of corrected CBCT are similar with plan CT images and dose calculations based on the modified CBCT show good agreement with plan CT.
Keywords:Tomography  X-ray computed  cone-beam  CT numbers regulate  Dose calculation  