首页 | 本学科首页   官方微博 | 高级检索  
     

10种介入诊疗程序中患者的辐射剂量调查
引用本文:张翼,宋少娟,曲桂莲,刘卫东,郭建军. 10种介入诊疗程序中患者的辐射剂量调查[J]. 中华放射医学与防护杂志, 2011, 31(4): 482-484. DOI: 10.3760/cma.j.issn.0254-5098.2011.04.029
作者姓名:张翼  宋少娟  曲桂莲  刘卫东  郭建军
作者单位:山东省医学影像学研究所,济南,250021
摘    要:
目的 调查研究介入诊疗程序中患者的受照剂量,评估其放射诊疗风险.方法 利用配置有符合IEC 60601-2标准的穿透型电离室的飞利浦Allura Xper FD20 DSA系统,收集记录10种介入诊疗程序共198例患者的剂量参数,估算出可供评估皮肤损伤的最高皮肤剂量及有效剂量.结果 累计透视时间范围为2.1~80.9 min,摄影帧数范围为15~678帧,剂量面积乘积范围为11~825 Gy·cm2,累计剂量范围为24~3374 mGy.有16例患者最高皮肤剂量超过1 Gy,79例患者有效剂量大于20 mSv.结论 有部分病例的最高皮肤剂量超过了皮肤损伤阈值,所以对患者的放射防护应给予足够的重枧.
Abstract:
Objective To investigate radiation dose to the patients undergoing interventional radiology and make radiation risk assessment.Methods Data was collected on 198 instances of 10 interventional radiology procedures by using Philips Allura Xper FD20 DSA, which was equipped with the transparent ionization chamber system in compliance with IEC 60601-2.Patient peak skin dose and effective dose were estimated.Results Cumulative fluoroscopy time was 2.1 - 80.9 min, and number of images monitored for PSD were above 1 Gy and 79 cases monitored for E were above 20 mSv.Conclusions Substantial number of cases exceeded the dose threshold for erythema.Due attention should be paid to radiation protection of patients.

关 键 词:介入放射学  照射剂量  辐射防护
收稿时间:2010-12-23

Radiation exposure to patients during 10 interventional radiology procedures
ZHANG Yi,SONG Shao-juan,QU Gui-lian,LIU Wei-dong and GUO Jian-jun. Radiation exposure to patients during 10 interventional radiology procedures[J]. Chinese Journal of Radiological Medicine and Protection, 2011, 31(4): 482-484. DOI: 10.3760/cma.j.issn.0254-5098.2011.04.029
Authors:ZHANG Yi  SONG Shao-juan  QU Gui-lian  LIU Wei-dong  GUO Jian-jun
Affiliation:Shandong Medical Imaging Research Institute, Jinan 250021, China;Shandong Medical Imaging Research Institute, Jinan 250021, China;Shandong Medical Imaging Research Institute, Jinan 250021, China;Shandong Medical Imaging Research Institute, Jinan 250021, China;Shandong Medical Imaging Research Institute, Jinan 250021, China
Abstract:
Objective To investigate radiation dose to the patients undergoing interventional radiology and make radiation risk assessment. Methods Data was collected on 198 instances of 10 interventional radiology procedures by using Philips Allura Xper FD20 DSA, which was equipped with the transparent ionization chamber system in compliance with IEC 60601-2.Patient peak skin dose and effective dose were estimated. Results Cumulative fluoroscopy time was 2.1-80.9 min, and number of images was 15-678 frames, and DAP was 11-825 Gy·cm2, and CD was 24-3374 mGy. A total of 16 cases monitored for PSD were above 1 Gy and 79 cases monitored for E were above 20 mSv. Conclusions Substantial number of cases exceeded the dose threshold for erythema. Due attention should be paid to radiation protection of patients.
Keywords:Interventional radiology  Radiation dose  Radiation protection