Evaluation of accuracy of computer-aided intraoperative positioning of endosseous oral implants in the edentulous mandible
Authors:Wanschitz Felix  Birkfellner Wolfgang  Watzinger Franz  Schopper Christian  Patruta Sanda  Kainberger Franz  Figl Michael  Kettenbach Joachim  Bergmann Helmar  Ewers Rolf
Affiliation:Department of Oral and Maxillofacial Surgery, University of Vienna, General Hospital Vienna, Austria. felix.wanschitz@chello.at
Abstract:
The overall accuracy of a novel surgical computer-aided navigation system for placement of endosseous implants was evaluated. Five dry cadaver mandibles were scanned using high resolution computed tomography (HRCT). The position of four interforaminal dental implants was planned on the computer screen and transferred to the cadaver mandibles using VISIT, a surgical navigation software developed at the Vienna General Hospital. The specimens were HRCT-scanned again to compare the position of the implants with the preoperative plan on reformatted slices after matching of the pre- and postoperative data sets using the mutual information technique. The overall accuracy was 0.96 +/- 0.72 mm (range 0.0-3.5 mm). No perforation of the mandibular cortex or damage to the mandibular canal occurred. We conclude that computer-aided implant surgery can reach a level of accuracy where further clinical developments are feasible.
Keywords:computer-aided surgery    implant dentistry    accuracy    mutual information matching