双源CT双能量心肌灌注成像的图像质量及辐射剂量研究
引用本文:齐燕,欧陕兴,钱民,彭光明,刘波,刘玉品,.双源CT双能量心肌灌注成像的图像质量及辐射剂量研究[J].放射学实践,2012,27(10):1095-1099.
作者姓名:齐燕  欧陕兴  钱民  彭光明  刘波  刘玉品  
作者单位:[1]广东省中医院放射科,广州510120 [2]广州军区总医院放射科,广州510010
摘    要:目的:探讨运用双源CT双能量技术行冠状动脉血管成像的图像质量和辐射剂量。方法:204例患者分为两组行双源CT冠状动脉血管造影检查,A组采用双能量扫描方案,B组采用传统单能量扫描方案,比较两组的图像质量及辐射剂量。结果:A、B两组冠状动脉图像质量评分分别为(4.70±0.72)分、(4.70±0.71)分,两组间差异无统计学意义(t=-1.000,P=0.318)。A、B两组图像左、右冠状动脉信噪比(SNR)分别为A组20.0±3.4和19.0±3.0、B组20.1±3.2和19.9±3.2,且两组间差异均无统计学意义(t=1.967,P=0.051;t=-0.233,P=0.816)。A、B两组的容积CT剂量指数(CTDIvol)、剂量长度乘积(DLP)、有效辐射剂量(ED)分别为(48.27±8.70)mGy、(54.10±8.77)mGy;(565.63±115.48)mGy.cm、(702.75±144.04)mGy.cm;(9.62±1.96)mSv、(11.95±2.45)mSv,A组的CTDIvol、DLP及ED明显低于B组,差异均有统计学意义(t=-4.764,P=0.000;t=-7.501,P=0.000;t=-7.501,P=0.000)。结论:双能量心肌灌注成像技术在不降低图像质量的情况下,能够显著降低患者的有效辐射剂量,有很高的临床应用价值。

关 键 词:体层摄影术  X线计算机  血管造影术  辐射剂量

Study of image quality and radiation dose with dual-energy technique in myocardial perfusion imaging using dual-source CT
Institution:QI Yan,OU Shan-xing,QIAN Min,et al.Department of Radiology,Guangdong Province Hospital of Chinese Traditional Medicine,Guangzhou 510120,P.R.China
Abstract:Objective:To investigate the image quality and radiation dose of dual-energy technique for myocardial perfusion imaging with dual source CT.Methods:204 patients undergone coronary CT angiography(CTA) were divided into two groups,group A with dual energy CT and group B with single-energy scanning protocol respectively.The radiation dose and image quality of these two groups were compared.Results:The image quality scores of coronary artery for group A and group B were 4.70±0.72 and 4.70±0.71,respectively,with no statistic difference(t=-1.000,P=0.318).The signal noise ratio(SNR) of left and right coronary artery in group A and group B were 20.0±3.4,19.0±3.0 and 20.1±3.2,19.9±3.2 respectively,with no statistic difference(t=-0.233,1.967,P=0.816,0.051).The CT dose index of volume(CTDIvol),dose length product(DLP) and effective dose(ED) were(48.27±8.70)mGy,(565.63±115.48)mGy·cm,(9.62±1.96)mSv respectively for group A and(54.10±8.77)mGy,(702.75±144.04)mGy·cm,(11.95±2.45)mSv respectively for group B.The CTDI vol and ED of group A were obviously lesser than that of group B,with significant statistic difference(t=-4.764,P=0.000;t=-7.501,P=0.000;t=-7.501,P=0.000).Conclusion:Using the dual-energy technique with dual source CT for myocardial perfusion imaging can obviously reduce the radiation dose while with no influence on image quality,which provides great potential in clinical application.
Keywords:Tomography  X-ray computed  Angiography  Radiation dose