首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of a Transcutaneous Oxygen and Carbon Dioxide Monitor in a Neonatal Intensive Care Department
Authors:PETER D. WIMBERLEY  PETER S. FREDERIKSEN  JESS WITT-HANSEN  STEEN G. MELBERG  BENT FRIIS-HANSEN
Affiliation:Departments of Neonatology, GN and Clinical Chemistry ML, Rigshospitalet, University of Copenhagen, Denmark
Abstract:
Transcutaneous-Po2; (tc-Po2 (tc-Pco2) at 44oC and transcutaneous-Pco2) at 38, 42, 43 and 44oC were measured with a currently available monitoring system (TCM222, Radiometer, Copenhagen) in 64 newborn infants with severe respiratory insufficiency during the first five days of life. Tc-Pco2 at all four temperatures correlated better with arterial blood-Pco2 (aB-Pco2)> than tc-Po2 with aB-Po2. However, the sensitivity and specificity of tc-Po2 and tc-Pco2 were similar with regard to maintaining aB-Po2, and aB-Pco2 within specified limits. Tc-Pco2 increased relatively with increasing electrode temperature by a factor which was similar to the anaerobic temperature coefficient of Po2, in blood. The coefficient of variation of duplicate measurements was 10 % for tc-Po2 and 5 % for tc-Pco2. Electrode drift after an average of 3 hours patient monitoring was 2%±6% (1SD) for tc-Pco2, and -3 %±6 % for tc-Pco2- We conclude that tc-Po2 and tc-Pco2 are a valuable supplement to arterial blood gas measurements, but the variable correlation with arterial blood gas values and the electrode drifts which may occur, mean that they cannot fully replace arterial sampling.
Keywords:Blood gases    arterial-Po2    orterial-Pco2    newborns    respiratory insufficiency    transcutaneous-Po2 transcutaneous-Pco2-