Single Capacitive Discharge Utilizing an Auxiliary Shock in the Coronary Venous System Reduces the Defibrillation Threshold
Authors:Paul Richard Roberts  Youhua Zhang  Shaowei Zhuan  Kent A. Mowrey  Don W. Wallick  Donald G. Hills  Tim Rider Betts  Stuart Allen  Jan Ewert  Todor N. Mazgalev  John Mark Morgan
Affiliation:(1) Wessex Cardiothoracic Center, Southampton, United Kingdom;(2) Cleveland Clinic Foundation, OH, USA
Abstract:Auxiliary shocks (AS) from electrodes sutured to the left ventricle (LV) prior to primary biphasic shocks (PS) have been shown to reduce defibrillation thresholds (DFT). Two capacitors are required to generate these waveforms. We investigate delivery of AS from one capacitor using a novel waveform. The epicardial surface of the LV is accessed transvenously via the middle cardiac vein (MCV) avoiding a thoracotomy. Methods: A defibrillation electrode was placed in the right ventricle (RV) and superior vena cava (SVC) in 12 pigs (37±2kg). A 50×1.8mm electrode was inserted in the MCV through a guide catheter. A can was placed in the left pectoral region. A monophasic AS (100mgrF, 1.5J) was delivered along one pathway before switching to deliver a biphasic waveform (40% tilt, 2ms phase 2) along another. DFTs (PS+AS) were assessed using a binary search. Two configurations not incorporating AS acted as controls. DFTs were compared using repeated measures analysis of variance. Results: DFTs of the four novel configurations (AS/PS) were: RVrarrCan/MCVrarrCan=14.9±3.7J, MCVrarrCan/RVrarrCan=17.2±5.7J, RVrarrSVC+Can/MCVrarrSVC+Can=13.4±4.6J, MCVrarrSVC+Can/RVrarrSVC+Can=17.1±5.9J. Delivering AS in the RV followed by PS in the MCV reduced the DFT (RVrarrCan (19.9±7.3 J, P<0.01) and RVrarrSVC+Can (19.2±6.0 J, P<0.05)). Conclusions: Delivering AS prior to PS in the MCV reduces the DFT by up to a third compared to conventional configurations of RVrarrCan and RVrarrSVC+Can. This is possible using only a single capacitor and an entirely transvenous approach to the LV.
Keywords:ventricular defibrillation  coronary veins  defibrillation threshold  capacitor  implantable cardioverter defibrillator
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号