首页 | 本学科首页   官方微博 | 高级检索  
     


Single-slice rebinning method for helical cone-beam CT
Authors:Noo F  Defrise M  Clackdoyle R
Affiliation:Université de Liège, Belgium. noo@doug.med.utah.edu
Abstract:
In this paper, we present reconstruction results from helical cone-beam CT data, obtained using a simple and fast algorithm, which we call the CB-SSRB algorithm. This algorithm combines the single-slice rebinning method of PET imaging with the weighting schemes of spiral CT algorithms. The reconstruction is approximate but can be performed using 2D multislice fan-beam filtered backprojection. The quality of the results is surprisingly good, and far exceeds what one might expect, even when the pitch of the helix is large. In particular, with this algorithm comparable quality is obtained using helical cone-beam data with a normalized pitch of 10 to that obtained using standard spiral CT reconstruction with a normalized pitch of 2.
Keywords:
本文献已被 PubMed 等数据库收录!