Leaflet escape in an Edwards TEKNA mitral prosthesis
Authors:Shinya Higuchi  Iwao Matsunaga  Kiyokazu Koga  Eiichi Teshima  Atsuhiro Nakashima  Ryuji Tominaga  Tsuyoshi Ito
Affiliation:1.Department of Cardiovascular Surgery,Shin-Yukuhashi Hospital,Yukuhashi,Japan;2.Department of Cardiovascular Surgery,Fukuoka Wajiro Hospital,Fukuoka,Japan
Abstract:
In almost every type of artificial valve, structural failure has been described. We are reporting on a case of a sudden leaflet escape of an Edwards TEKNA mitral valve prosthesis 12 years after implantation. The patient had a sudden onset of dyspnea and severe pulmonary edema with subsequent cardiogenic shock. An emergency mitral replacement was successfully performed. A multi-detector computed tomography scanning and three-dimensional imaging showed two fragments that had embolized in the terminal aorta and the left common iliac artery. The patient presented visual field abnormality, and postoperative head computed tomography showed watershed cerebral infarction. The escaped leaflet that fractured transversely was removed, following the patient’s recovery, during cardiac surgery.