7种半胱氨酸蛋白酶抑制剂C检测系统的分析性能评价
引用本文:安崇文,李海霞.7种半胱氨酸蛋白酶抑制剂C检测系统的分析性能评价[J].检验医学,2013(12):1106-1111.
作者姓名:安崇文  李海霞
作者单位:北京大学第一医院检验科,北京100034
摘    要:目的评估7种半胱氨酸蛋白酶抑制剂C(CysC)检测系统的分析性能。方法应用美国临床实验室标准化协会(CLSI)系列文件评估5个颗粒增强免疫透射比浊(PETIA)检测系统和1个均相溶胶颗粒免疫法(SPIA)检测系统的精密度、正确度、空白限(LoB)、抗干扰性以及与SIEMENSBNⅡ颗粒增强免疫散射比浊(PE—NIA)检测系统(简称BNⅡ系统)的相关性和偏差。结果6个检测系统(以A—F表示)和BN1I系统在CysC浓度为0.8~5.0mg/L时的总批内变异系数(CV)均〈3.75%,总批间CV除E系统外均〈5%;正确度验证显示A—E系统和BNⅡ系统测定ERM—DA471的绝对偏倚分别为0.00、0.00、1.01、-0.01、-0.50和-0.35mg/L,相对偏倚分别为0%、0%、18.43%、-0.18%、-9.12%和-6.39%;测定CAP室间质评物显示仅D系统和BN1I系统较为理想。A~F系统及BNⅡ系统的LoB分别为0.00、0.00、0.31、0.01、0.10、0.06和〈0.05mg/L。干扰分析显示Hb≤13g/L、TG≤28mmol/L时,对A、B系统及BNⅡ系统无明显影响(干扰〈±10%),其他系统均有不同程度干扰。相关分析显示A~F系统与BN1I系统相关性较好相关系数(r)均〉0.975,P〈0.01]。偏差分析显示,A~F系统与BNII系统的平均偏差分别为-0.02、-0.10、-0.31、0.05、0.04、0.36mg/L。结论CysC各检测系统测定有证参考物质ERM—DA471的最大偏倚可达到lmg/L,最大相对偏倚为18.43%;其中PENIA在测定ERM—DA471和CAP室间质评物时结果低于PETIA;部分检测系统分析性能存在不足。

关 键 词:半胱氨酸蛋白酶抑制剂C  颗粒增强免疫透射比浊检测系统  颗粒增强免疫散射比浊检测系统    相溶胶颗粒免疫检测系统

Performance evaluation of seven cystatin C analysis systems
AN Chongwen,LI Haixia.Performance evaluation of seven cystatin C analysis systems[J].Laboratory Medicine,2013(12):1106-1111.
Authors:AN Chongwen  LI Haixia
Institution:. (Department of Clinical Laboratory, Peking University First Hospital, Beijing 100034, China)
Abstract:Objective To evaluate the analysis performance for 7 cystatin C ( Cys C ) analysis systems. Methods A total of 5 particle-enhanced turbidimetric immunoassay (PETIA) systems and 1 sol particle immunoassay (SPIA) system were evaluated according to the documents of the Clinical and Laboratory Standards Institute (CLSI) , and their correlation with particle-enhanced nephelometric immunoassay (PENIA) system from the SIEMENS BN ]I system (BN 1I system) , precision, accuracy, limit of blank (LoB) and anti-interference ability were evaluated. Results For 6 systems (A-F) and BN 1I system, the total within-run coefficients of variation (CV) were all 〈 3.75% , and except E, the total between-run CV were all 〈 5% , when Cys C concentration was 0.8-5.0 mg/L. For A-E and BN 1I systems, accuracy verification showed that the absolute biases for measuring ERM-DA471 were 0.00,0.00,1.01, - 0.01, - 0.50 and -0.35 mg/L respectively,and the relative biases were 0% ,0% ,18.43%, -0.18%, -9. 12% and -6.39% respectively. The measurement of CAP external quality assessment materials showed that D and BN II systems had desirable results. The LoB were 0. 00,0. 00,0. 31,0. 01,0. 10,0.06 and 〈 0.05 mg/L for A-F and BN It systems, respectively. When Hb ~〈 13 g/L and TG ~〈 28 mmol/L, A, B and BN II systems were not significantly influenced (interference 〈 _+ 10% ). Other systems were influenced with different levels. The A-F systems correlated well with BNII system, and the correlation coefficients (r) were all 〉 0. 975 (P 〈 0.01 ). Compared with BN II system,the average deviations were -0. 02, -0. 10, -0. 31,0. 05,0. 04 and 0. 36 mg/L for A-F systems, respectively. Conclusions The maximum bias can reach 1 mg/L when measuring the reference materials for Cys C analysis systems, and the maximum relative bias is 18.43%. The results of PENIA system are lower than those of PETIA system when measuring ERM-DA471 and CAP external quality assessment materials. In a part of analysis systems, analysis performance is not good.
Keywords:Cystatin C  Particle-enhanced turbidimetric immunoassay system  Particle-enhanced nephelometric immunoassay system  Sol particle immunoassay system