High‐frequency ventilation with the Dräger Babylog 8000plus: measuring the delivered frequency
Authors:KI Wheeler  GM Schmölzer  CJ Morley  PG Davis
Affiliation:1. Neonatal Services, Royal Women’s Hospital (RWH), Melbourne, Australia;2. The Ritchie Center, Monash Institute for Medical Research, Monash University, Melbourne, Australia;3. Murdoch Childrens Research Institute, Royal Children’s Hospital, Melbourne, Australia;4. Division of Neonatology, Department of Paediatrics, Medical University, Graz, Austria;5. Department of Obstetrics and Gynaecology, University of Melbourne, Melbourne, Australia
Abstract:
Background: Ventilator frequency is one of the determinants of tidal volume delivery during high‐frequency ventilation. Clinicians increasingly use data on ventilator displays to inform their decisions. Aim: To measure the frequencies delivered by the Dräger Babylog 8000plus ventilator when used in high‐frequency mode. Methods: Ventilator waveforms using a test lung were recorded at the full range of settings 5–20 Hz using Spectra software at 1000 Hz. The changes in frequency produced by a 1‐ Hz change in set frequency were calculated. Actual and displayed frequencies were compared. Results: For settings up to 12 Hz, median (range) difference between set and delivered frequencies was 0 (?0.4 to +0.1) Hz. Above 12 Hz, delivered frequency varied by ?0.3 (?1.9 to +0.3) Hz. For 1‐ Hz changes in frequency settings, in the range 5–12 Hz, 1‐ Hz changes produced a change in delivered frequency of 1.0 (0.6–1.4) Hz. Above 12 Hz, the corresponding changes were 0.7 (0–2.9) Hz. The ventilator displays the set frequency during operation rather than the delivered frequency. Conclusion: At 12 Hz and below, the differences between set and delivered frequencies were relatively small compared with those at 13 Hz and higher. Above 13 Hz, the difference between set and delivered frequencies was up to 2.9 Hz. Some frequency setting changes did not result in a change in delivered frequency.
Keywords:High‐Frequency Ventilation  Infant  Intensive Care  Mechanical  Neonatal  Newborn  Respiratory Function Monitoring  Ventilators
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号