Radiosensitive functional dye: clinical application for estimation of patient skin dose |
| |
Authors: | Suzuki Shigeru Furui Shigeru Kohtake Hiroshi Takeshita Tohru Suzuki Masatoshi Kozuma Ken Yamamoto Yoshito Isshiki Takaaki |
| |
Affiliation: | Department of Radiology, Teikyo University School of Medicine, 2-11-1 Kaga, Itabashi-ku, Tokyo 173-8605, Japan. s-suzuki@med.teikyo-u.ac.jp |
| |
Abstract: | Institutional review board approval and informed patient consent were obtained. The purpose of the study was to prospectively evaluate the use of radiosensitive indicators to estimate patient entrance skin dose (ESD). Forty-six patients wore a jacket with 48 or 52 indicators adhered to the back during percutaneous coronary interventions; they had eight additional indicators on their upper arms. The patients' ESDs were calculated according to the change in color of the indicators. There were good correlations between the ESDs estimated by using color measurements performed with an optical instrument and those estimated at visual observation (P < .001) and between the ESDs estimated by using a thermoluminescent dosimeter and those estimated by using color measurements (P < .001). The radiosensitive indicator method seems to be useful for estimating ESDs and their distribution during percutaneous coronary intervention; however, visual observation is reliable for estimating doses of up to 5 Gy only. (c) RSNA, 2006. |
| |