A simple technique for transferring excised patches of membrane to different solutions for single channel measurements |
| |
Authors: | Nino Quartararo Peter H. Barry |
| |
Affiliation: | (1) School of Physiology and Pharmacology, University of New South Wales, Sydney, Australia |
| |
Abstract: | A technical problem associated with the patch clamp technique has been the changing of solutions bathing the membrane patch. The simple technique described here solves this problem by means of a movable polythene sleeve placed on the shaft of the patch clamp pipette. The sleeve is initially placed so that the tip of the pipette is exposed. A gigaohm seal is formed using standard techniques. The patch is then excised and the sleeve is slipped down a few mm past the end of the tip of the pipette. When the pipette and sleeve is now removed from the solution, a small drop of solution covering the membrane patch is held in place at the end of the sleeve by surface tension. The pipette is then easily transferred to a different solution without passing the membrane patch through the air-water interface. The sleeve is then simply pulled back up the pipette shaft to expose the membrane patch to the new solution. |
| |
Keywords: | patch clamp single channels excised patches solution changes |