首页 | 本学科首页   官方微博 | 高级检索  
     


A simple technique for transferring excised patches of membrane to different solutions for single channel measurements
Authors:Nino Quartararo  Peter H. Barry
Affiliation:(1) School of Physiology and Pharmacology, University of New South Wales, Sydney, Australia
Abstract:
A technical problem associated with the patch clamp technique has been the changing of solutions bathing the membrane patch. The simple technique described here solves this problem by means of a movable polythene sleeve placed on the shaft of the patch clamp pipette. The sleeve is initially placed so that the tip of the pipette is exposed. A gigaohm seal is formed using standard techniques. The patch is then excised and the sleeve is slipped down a few mm past the end of the tip of the pipette. When the pipette and sleeve is now removed from the solution, a small drop of solution covering the membrane patch is held in place at the end of the sleeve by surface tension. The pipette is then easily transferred to a different solution without passing the membrane patch through the air-water interface. The sleeve is then simply pulled back up the pipette shaft to expose the membrane patch to the new solution.
Keywords:patch clamp  single channels  excised patches  solution changes
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏