The first case of “locked‐in leaflet” after transcatheter aortic valve replacement in a patient with bicuspid aortic stenosis |
| |
Authors: | Nao Shibata MD Kensuke Matsumoto MD PhD Takayoshi Toba MD Shumpei Mori MD PhD Shinsuke Shimoyama MD Hidekazu Tanaka MD PhD Toshiro Shinke MD PhD Ken‐ichi Hirata MD PhD |
| |
Affiliation: | 1. Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine, Kobe, Japan;2. Department of Radiology, Kobe University Graduate School of Medicine, Kobe, Japan |
| |
Abstract: | This case report is about an 85‐year‐old woman with bicuspid aortic stenosis (AS). Although preoperative multimodality imaging showed challenging anatomical aspects, transcatheter aortic valve replacement (TAVR) was selected to be performed as a less invasive alternative treatment approach, owing to her comorbidity. Postoperative transesophageal echocardiography and multidetector‐row computed tomography revealed the presence of “locked‐in leaflet” caused by stent distortion due to pinching by calcified native leaflets, with nodular calcification preventing the full expansion of the valve, which resulted in moderate perivalvular leakage. This is the first reported case of bicuspid AS treated with TAVR that eventually resulted in “locked‐in leaflet.” |
| |
Keywords: | aortic stenosis bicuspid aortic valve locked‐in leaflet multidetector‐row computed tomography transcatheter aortic valve replacement |
|
|