彩色速度成像及脉冲多普勒测量正常人颈动脉流速对照
引用本文:张琦,李寒. 彩色速度成像及脉冲多普勒测量正常人颈动脉流速对照[J]. 中国临床医学影像杂志, 1999, 10(3): 161-163
作者姓名:张琦  李寒
作者单位:第二军医大学长海医院超声科(张琦),一汽吉林轻型车厂医院(李寒)
摘    要:目的:探讨彩色速度成像(CVI)测量血流速度的准确性。方法:对60位正常人分别用CVI和脉冲多普勒方法测量左右颈动脉的最大血流速度(Vmax)和阻抗系数(RI)并进行统计学处理。结果:两种方法测量女性左侧颈动脉Vmax分别为7744±1851cm/s、8360±1990cm/s(r=0959),RI分别为0733±0054、0719±0047(r=0904);右侧颈动脉Vmax分别为7395±1918cm/s、7949±1822cm/s(r=0955),RI分别为0736±0058、0734±0061(r=0927);男性左侧颈动脉Vmax分别为9417±2435cm/s、9776±2408cm/s(r=0971),RI分别为0759±0064、0743±0057(r=0910);右侧颈动脉Vmax分别为9196±2389cm/s、9796±2497cm/s(r=0954),RI分别为0789±0052、0783±0055(r=0902)。结论:两种方法对血流Vmax及RI的测量均相关很好(p<0001),证实了CVI技术测量

关 键 词:彩色速度成像 脉冲多普勒 颈动脉

Comparison of flow velocity measurement of the carotid arteries using color velocity imaging and pulsed Doppler ultrasound in healthy persons
Zhang Qi,Li Han. Comparison of flow velocity measurement of the carotid arteries using color velocity imaging and pulsed Doppler ultrasound in healthy persons[J]. Journal of China Clinic Medical Imaging, 1999, 10(3): 161-163
Authors:Zhang Qi  Li Han
Affiliation:Zhang Qi,Li HanDepartment of Ultrasound,the Second Military Medical University,Changhai Hospital,Shanghai,200433
Abstract:Objective:In this paper we studied the accuracy of flow velocity measurement by Color velocity Imaging(CVI)Method:We measured Vmax and RI of left and right carotid arteries in 60 normal persons using CVI and pulsed Doppler ultrasound and analysed by statistics.Result:The measurement results of CVI and pulsed Doppler ultrasound showed:the Vmax was 77441851cm/s83601990cm/s(r=0959) and RI was 0733005407190047(r=0904) detected in female left carotid arteries;the Vmax was 73951918cm/s79491822cm/s(r=0955) and RI was 0736005807340061(r=0927) detected in female right carotid arteries;the Vmax was 94172435cm/s97762408cm/s(r=0971)and RI was 0759006407430057(r=0910) detected in male left carotid arteries;the Vmax was 91962389cm/s97962497cm/s(r=0954)and RI was 0789005207830055(r=0902) detected in male right carotid arteries. Conclusion:The all correlations were statistically significant(p<0001).The results verified the accuracy of flow velocity measureing by CVI.
Keywords:Color velocity imagungPulsed Doppler ultrasoundCarotid artery  