快速多时相减影CE-MRA对头颈部TOF-MRA伪影的矫正
引用本文:韩鸿宾,沙琳,刘洪东,许峰,边杰,刘溢,杨玉杰.快速多时相减影CE-MRA对头颈部TOF-MRA伪影的矫正[J].中国医学影像技术,2003,19(5):632-635.
作者姓名:韩鸿宾  沙琳  刘洪东  许峰  边杰  刘溢  杨玉杰
作者单位:1. 北京大学第三医院放射科,北京,100083
2. 大连医科大学附属第二医院放射科
3. 鞍山市中心医院放射科
4. 北京大学第三医院医工处,北京,100083
5. 北京大学第三医院超声科,北京,100083
基金项目:北京大学"985"临床多中心前瞻性研究资助项目.
摘    要:目的 通过总结TOF-MRA(time of flight MRA)常见伪影的表现,分析其形成原因以及研究CE-MRA(contrast enhanced MRA)对其的矫正作用。方法 同时进行头颈部TOF-MRA与CE-MRA检查患者共217例,以手术或DSA结果为金标准,对TOF-MRA出现的各种伪影与CE-MRA图像进行对比分析。结果 在217例患者中,TOF-MRA出现伪影266处,其中狭窄后湍流伪影25处,流动分离67处,流动移位37处,边缘锯齿伪影72处,威氏暗带伪影65处,CE-MRA对TOF-MRA中相对应部位的伪影均进行了矫正。结论 头颈部TOF-MRA中伪影是常见的现象,需要正确认识这些伪影,提高诊断的正确率;TOF-MRA出现的伪影可以通过CE-MRA进行矫正。

关 键 词:磁共振  血管成像  造影剂  伪影
文章编号:1003-3289(2003)05-0632-04
收稿时间:2003/1/10 0:00:00
修稿时间:2003年1月10日

Correction of the Artifacts in Cranial and Cervical TOF-MRA by Rapid Multiple-Phase CE-MRA
HAN Hong-bin,SHA Lin,LIOU Hong-dong,XU Feng,BIAN Jie,LIU Yi and YANG Yu-jie.Correction of the Artifacts in Cranial and Cervical TOF-MRA by Rapid Multiple-Phase CE-MRA[J].Chinese Journal of Medical Imaging Technology,2003,19(5):632-635.
Authors:HAN Hong-bin  SHA Lin  LIOU Hong-dong  XU Feng  BIAN Jie  LIU Yi and YANG Yu-jie
Institution:Departement of Radiology,Peking Universrity Third Hospital,Beijing 100083,China;Departement of Radiology,Peking Universrity Third Hospital,Beijing 100083,China
Abstract:Objective To summarize the appearances of the artifacts in cranial and cervical TOF-MRA. To analyze the causes and to study the artifact correction of contrast enhanced MRA (CE-MRA). Methods TOF-MRA and CE-MRA examinations were performed in 217 patients simultaneously. Artifacts appeared in TOF-MRA and the corresponding images of CE-MRA were summarized and analyzed. The results were compared with the findings on DSA or in surgery. Results Among the 217 cases, there were 266 artifacts in TOF-MRA . The artifacts included turbulent behind stenosis (25), mobile separations (67), mobile shifts (37), margin zigzag artifacts (72) and dark band (65). CE-MRA corrected all the artifacts appeared in TOF-MRA. Conclusion Artifacts,which are common,could be corrected by CE-MRA.Eavaluating them correctly are needed to improve the diagnostic accuracy.
Keywords:Magnetic resonance  Vascular imaging  Contrast medium  Artifact