Compression forces generated by Mini bone screws--a comparative study done on bone model |
| |
Authors: | Adla Deepthi Nandan Kitsis C Miles A W |
| |
Affiliation: | Department of Orthopaedic Engineering, University of Cardiff, Leicester General Hospital, Leicester, UK. deepthinandan@doctors.org.uk |
| |
Abstract: | The compressive forces generated by the AO/ASIF 3.0 mm cannulated cancellous and 2.0 mm cortical screws, Mini-Acutrak and Herbert/Whipple small bone cannulated screws were measured in the laboratory with the use of simulated cancellous bone and a load cell washer as a means of quantifying their fixation capabilities. The Herbert/Whipple screw and the Mini-Acutrak screw were found to have nearly identical compression capabilities and provided more compression than the cortical screw. The AO/ASIF cannulated screw when used with a support screw demonstrated a compressive capacity twice that of the 2.0 mm cortical screw and higher than the headless Mini-Acutrak and Herbert/Whipple screws. The Mini-Acutrak screw produced about 70% of compression of the cancellous screw in spite of having a diameter almost half that of the cancellous screw. The Herbert/Whipple screw in spite of its larger size compared to the Mini-Acutrak produced almost the same amount of compression. |
| |
Keywords: | Mini bone screws Load cell Compression forces Simulated bone model Experimental study |