首页 | 本学科首页   官方微博 | 高级检索  
     


Cone beam collimation for single photon emission computed tomography: analysis, simulation, and image reconstruction using filtered backprojection
Authors:R J Jaszczak  C E Floyd  S H Manglos  K L Greer  R E Coleman
Abstract:This paper presents an analysis of two cone beam configurations (having focal lengths of 40 and 60 cm) for the acquisition of single photon emission computed tomography (SPECT) projection data. A three-dimensional filtered backprojection algorithm is used to reconstruct SPECT images of cone beam projection data obtained using Monte Carlo simulations. The mathematical analysis resulted in on-axis point source sensitivities (calculated for a distance of 15 cm from the collimator surface) for cone beam configurations that were 1.4-3 times the sensitivities of parallel-hole and fan beam geometries having similar geometric resolutions. Cone beam collimation offers the potential for improved sensitivity for SPECT devices using large-field-of-view scintillation cameras.
Keywords: