An evaluation of two conducted electrical weapons using a swine comparative cardiac safety model |
| |
Authors: | Donald M. Dawes Jeffrey D. Ho Johanna C. Moore Andrew P. Laudenbach Robert F. Reardon James R. Miner |
| |
Affiliation: | 1. Lompoc Valley Medical Center, Lompoc, CA, USA 2. Santa Barbara Police Department, Santa Barbara, CA, USA 3. Department of Emergency Medicine, Hennepin County Medical Center, 701 Park Avenue South, Minneapolis, MN, 55415, USA 4. Meeker County Sheriff’s Office, Litchfield, MN, USA
|
| |
Abstract: | Arrest-related deaths proximate to the use of a conducted electrical weapon (CEW) continue to generate controversy despite a better understanding of the multi-factorial nature of many of these deaths. With the rapid adoption of this technology by law enforcement, and the proliferation of companies entering the marketplace, it is important to have a method to assess the relative safety of these weapons. We had previously developed a model to assess the relative cardiac safety of CEWs. In this study, we use this model to compare the TASER X2 and the Karbon Arms MPID. Our results suggest that the TASER X2 may have an improved cardiac safety margin over the Karbon Arms MPID as determined by a smaller area of cardiac pacing on the anterior chest in our model. This model seems to offer a reproducible means of comparing the cardiac effects of CEWs. |
| |