首页 | 本学科首页   官方微博 | 高级检索  
     


Removal of a well-fixed trabecular metal monoblock tibial component
Authors:Klein Gregg R  Levine Harlan B  Hartzband Mark A
Affiliation:Hackensack University Medical Center, Hackensack, New Jersey 07652, USA.
Abstract:
The use of porous tantalum (Trabecular Metal; Zimmer, Inc, Warsaw, Ind) in hip and knee reconstruction has become increasingly popular over the past few years. Widespread clinical use of porous tantalum tibial components for primary total knee arthroplasty has been tempered in part by the perceived difficulty in removing this implant after bone ingrowth has occurred. We present an easy, reproducible, and inexpensive technique for removal of a well-fixed Trabecular Metal Monoblock Tibial Component (Zimmer), which has been used in 4 revision knees. This technique does not require the use of any specialized equipment and results in the production of minimal metallic debris.
Keywords:trabecular metal   tantalum   monoblock knee   explant   knee revision
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号