High-energy electron and photon therapy to the parotid bed: radiation dose perturbations with a titanium mandibular implant
Authors:Menard C M  Bews J  Skoracki R I  Chowdhury A D
Affiliation:Department of Radiation Oncology, Manitoba Cancer Treatment and Research Foundation, Winnipeg, Canada. umskorac@cc.umanitoba.ca
Abstract:
Adjuvant radiation therapy to the parotid bed is commonly administered following surgical resection using either a pair of angled wedged photon beams or an ipsilateral mixed-beam portal of electrons and photons. The present study seeks to determine the optimal parotid bed treatment technique in the presence of a titanium mandibular implant by investigating perturbations in the dose distribution deep to this implant for a 15-MeV electron beam and a 6-MV photon beam. A titanium mandibular plate was embedded in a tissue-equivalent phantom, and irradiated with 15-MeV electrons, and 6 MV photons. Radiation doses behind the plate were measured with both thermoluminescent dosimeters and radiographic film. With 15-MeV electrons, there is a clinically significant decrease in the dose beyond the titanium plate, which is most important at 5-mm and 10-mm depths (18-27%). With 6-MV photons the dose at the deep interface of titanium and tissue is reduced by between 15 and 18%, but rapidly drops to < 5% at a depth of 5 mm. In adjuvant treatment to the parotid bed, when the clinical target volume includes tissue positioned deep to a titanium implant, significant underdosage occurs with ipsilateral beam arrangements, especially when electrons are used.