首页 | 本学科首页   官方微博 | 高级检索  
     


Biomechanical evaluation of maxillary Lefort Ι fracture with bioabsorbable osteosynthesis internal fixation
Authors:Wei Wu  Jiang Zhou  Chong‐Tao Xu  Jie Zhang  Yan‐Jiao Jin  Geng‐Lin Sun
Affiliation:Department of stomatology, Tianjin medical university general hospital, , Tianjin, China
Abstract:
The aim of this study was to apply biomechanical analysis model to evaluate the effects of bioabsorbable internal fixation devices on maxillary Lefort Ι fracture. CT scan technology and the finite element software (ansys ) were used to establish three‐dimensional finite element models of five resorbable internal fixation devices in maxillary Lefort Ι fractures. We used the model to calculate the stress of the upper jaw and internal fixation. We further analyzed the stability of fixation under four occlusions. The fixation using two bioabsorbable plates was not stable. The zygomaticomaxillary pillars fixation is more stable than other fixations. The stability of fracture fixation was influenced with the molar occlusion. The current study developed a functional three‐dimensional finite element model of bioabsorbable internal fixation and compared the stability of five fixation methods for maxillary Lefort Ι fractures. The results would facilitate the application of bioabsorbable materials in dental clinic.
Keywords:bioabsorbable materials  finite element  internal fixation  maxillary fractures
设为首页 | 免责声明 | 关于勤云 | 加入收藏