基于增强现实在髋关节术前定位中的动物实验研究
引用本文:邹昆 王淇,于东方 丁冉 张启栋,郭万首 王卫国.基于增强现实在髋关节术前定位中的动物实验研究[J].生物骨科材料与临床研究,2021,18(2):1-5.
作者姓名:邹昆 王淇  于东方 丁冉 张启栋  郭万首 王卫国
作者单位:北京中医药大学;北京大学;北京航空航天大学;中日友好医院骨科
基金项目:国家重点研发计划课题(2017YFC0108102);北京市科技计划(首特)课题(Z171100001017209)。
摘    要:目的利用基于X线叠加技术的增强现实(AR)技术在比格犬上进行髋关节术前精确定位,并评估定位效果。方法取20只比格犬,采用基于X线的二维图形叠加技术的AR技术进行术前定位,通过术中C型臂验证与术中所见目标的实际位置对比,评估该方法的定位准确度、结果可重复性,计算总定位平均用时和C型臂投照次数,记录实验过程中不良事件。结果正位定位误差(1.05±0.81) mm,侧位定位误差(1.29±1.26) mm,但是正位的Lerror (反映了该实验结果的可重复性)为(64.89±99.90) %,侧位片的Lerror为(108.61±156.93)%,总定位平均时长(18.25±9.68) min,C型臂平均投照次数(6.35±2.33)次,定位准确度符合预期,实验过程中无不良事件发生。结论采用基于X线叠加技术的AR技术进行髋关节术前定位安全、精准。然而,该系统的定位准确性仍需更多的临床实践进一步明确。

关 键 词:增强现实  髋关节  动物研究  图像叠加技术  定位

Assessment of augmented reality performance for preoperative positioning in beagles' hip joint
Zou Kun,Wang Qi,Yu Dongfang,Ding Ran,Zhang Qidong,Guo Wanshou,Wang Weiguo..Assessment of augmented reality performance for preoperative positioning in beagles' hip joint[J].Orthopaedic Biomechanics Materials and Clinical Study,2021,18(2):1-5.
Authors:Zou Kun  Wang Qi  Yu Dongfang  Ding Ran  Zhang Qidong  Guo Wanshou  Wang Weiguo
Institution:(Beying University of Chinese Medicine,Beijing,100029;Peking University Beijing,100191;Beijing University of Aeronautics and Astronautics,Bering,100191;Department of Orthopaedic Surgery,Beijing Key Lab Immunes-Mediated Inflammatory Diseases,China-Japan Friendship Hospital,Bering,100029,China)
Abstract:Objective To prospectively assess overlay technology in providing accurate and efficient targeting for X-ray imaging-guided beagles’ hip puncture.Methods Twenty beagle dogs were selected,and AR technology based on X-ray two-dimensional graphics overlay technology was used for preoperative positioning.The intraoperative C-arm was used to verify the actual position of the target seen during the operation.The customized method was used to evaluate the system’s positioning accuracy,measurement repeatability,total positioning average time and average C-arm shot times,and adverse events during the experiment were recorded.Results The frontal positioning error was(1.05±0.81) mm,the lateral positioning error was(1.29±1.26) mm,the frontal Lerror(the score for measurement repeatability) was(64.89±99.90) %,the lateral was(108.61 ±156.93) %,and the positioning is accurate with expectations,the average number of shots of the C-arm was(6.35±2.33) times.The average positioning time was(18.25±9.68) min,and no adverse event occurred during the experiment.Conclusion The use of AR based on X-ray imaging-guided technology is safe and accurate.However,the positioning accuracy of the system still needs more clinical practice to further clarify.
Keywords:Augmented reality  Hip  Animal testing  Image overlay technology  Positioning