Comparative clinical study on deformity correction accuracy of different external fixators
Authors:Ilker Eren  Levent Eralp  Mehmet Kocaoglu
Affiliation:1. Orthopaedics and Traumatology Department, Erzurum Area Training and Research Hospital, Erzurum Bolge Egitim ve Arastirma Hastanesi, Palandoken, Erzurum, Turkey
2. Istanbul Faculty of Medicine, Orthopaedics and Traumatology Department, Istanbul University, Istanbul, Turkey
3. Orthopaedics and Traumatology Department, Memorial Hospital, Istanbul, Turkey
Abstract:

Purpose

In this study, the correction accuracy of Smart Correction spatial fixators and of Ilizarov-type external fixators are compared in terms of deformity complexity.

Methods

Seventy-seven (40 male, 37 female) bone segments of 57 patients treated with a Smart Correction device were compared with 94 (51 male, 43 female) segments of 68 patients treated with an Ilizarov fixator. Mean age of the Smart Correction group was 20.69?±?12.94 years, and or the Ilizarov group 22.45?±?12.18 years. Patients were categorised according to limb lengthening and the number of deformity planes.

Results

A longer correction period was found with Ilizarov (66.53?±?47.7 days) compared with Smart Correction (49.05?±?35.6 days) devices. The bone healing index of the Ilizarov group was significantly better compared with the spatial group. Residual deformity after treatment was significantly lower with the Smart Correction device; however, this relationship could not be shown between subgroups. Although there was no significant difference between subgroups, mean residual deformity was higher with the increasing number of planes of the deformity.

Conclusions

The Smart Correction fixator is an accurate device that allows ease of application and planning. It demonstrates higher accuracy for correcting deformities compared with an Ilizarov external fixator. With an increasing number of planes, the difference between the two devices becomes even more pronounced. The relationship between the complexity of the deformity and residual deformity may possibly be significantly greater in favour of the Smart Correction fixator in a study with a larger sample size.