首页 | 本学科首页   官方微博 | 高级检索  
     


Percutaneous insertion of the Kimray-Greenfield filter: technical considerations and problems
Authors:Pais, SO   Mirvis, SE   De Orchis, DF
Affiliation:Department of Diagnostic Radiology, University of Maryland Medical System/Hospital, Baltimore 21201.
Abstract:Kimray-Greenfield filters were inserted percutaneously into the inferior vena cava (IVC) in 57 patients. Thirty-six were placed from the right femoral vein, 14 from the left femoral vein, and seven from the right internal jugular vein. There were no deaths or major complications and only six minor complications. Inferior vena cavography was done before filter insertion in all cases. Cavography is vital to determine feasibility of filter insertion, route of insertion, and filter location; pertinent findings include caval size, presence or absence of clot in the IVC or iliac veins, and position of the renal veins. The guide wire provided with the standard filter introduction set has a tendency to catch on the filter as the wire is withdrawn. A stiff wire with a straight, tapered, floppy tip was substituted. The femoral approach is preferred when it is feasible. Though there was only one known occurrence of femoral vein thrombosis at the filter insertion site, other cases may have occurred and may not have been detected. If the frequency of this complication proves to be significant, the preferred route for filter insertion may have to be reconsidered.
Keywords:
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏