首页 | 本学科首页   官方微博 | 高级检索  
     


The dynamic range of TMJ sounds
Authors:Widmalm S E  Djurdjanovic D  McKay D C
Affiliation:Department of Biologic Sciences, School of Dentistry, University of Michigan, USA. sew@umich.edu
Abstract:
It is of clinical interest to record the amplitudes of temporomandibular joint (TMJ) sounds. The aim was to test the hypothesis that sealing the meatus, when placing a microphone in the ear canal affects such recording by increasing the sound pressure level (SPL). Bilateral recordings of 249 TMJ clickings were made from three subjects, using sampling rates of 48 or 96 kHz and 24 bits A/D conversion, with and without the ear canals sealed by Silicone putty. The peak-to-peak equivalent sound pressure level (peSPL) was higher (P < 0.001) when the ear canal was sealed (range of mean differences was 8.3-24.9 dB peSPL). This means that the signal to noise ratio can be improved by sealing the meatus because the electronic noise level is not increased. Most important is that the dynamic range of the clicking sounds was 62 dB that is larger than the effective dynamic range of a 16 bits sound card. Future studies are needed to establish normative peSPL values. However, cards with at least 24 bits A/D conversion will be required, especially in patients with suspected disc displacement with reduction, where the difference in loudness between opening and closing clicking often is large.
Keywords:temporomandibular joint    clicking    sound pressure level    signal processing    dynamic range    disc displacement    signal to noise ratio