Digital imaging of the chest in newborns: influence of radiation exposure on image quality
Authors:U. Bick  R. Hentschel  M. Fiebich  C. Müller-Leisse  N. Ross  W. Wiesmann  H. Lenzen  H.-J. v. Lengerke  P. E. Peters
Affiliation:(1) Department of Radiology, University of Münster, Münster, Germany;(2) Department of Paediatrics, University of Münster, Münster, Germany;(3) Department of Radiology, University of Aachen, Aachen, Germany;(4) Present address: Department of Radiology, University of Chicago, 5841 South Maryland Avenue, 60637 Chicago, IL, USA
Abstract:
In order to evaluate the influence of radiation exposure on image quality in digital storage phosphor radiography, 200 digital storage phosphor chest radiographs, obtained on a neonatal Intensive Care Unit, and the same number of conventional screen-film radiographs (speed 250) were assessed for the visibility of anatomical structures and catheters. The same exposure parameters were used in both groups. Normal variations of radiation exposure under free exposure conditions were estimated in all digital images using a formula calculating radiation dose in the screen-plane from image sensitivity, latitude and average grey value of the right lung. There was already a significant (P < 0.001) decline in image quality in the digital images with a 30–50% reduction in radiation exposure, which was most severe for structures such as trachea, retrocardiac space, lung texture and low-contrast catheters. Compared with optimally exposed conventional images, only those digital images with a slightly higher than normal dose had an equivalent image quality.Correspondence to: U. Bick
Keywords:Chest  Digital storage phosphor radiography  Intensive care  Neonates  Radiation dose
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号