A comparative study of plateletpheresis using Baxter Autopheresis C and Haemonetics PCS Plus
Authors:J D Walton  E A Caffrey  J P Allain
Institution:East Anglian Blood Transfusion Centre and Tissue Bank;*University of Cambridge Division of Transfusion Medicine
Abstract:SUMMARY. This study compared plateletpheresis on the Haemonetics PCS Plus (PCS Plus) and the Baxter Autopheresis C (Auto C) using the same 100 selected donors. The number of packs meeting UK BTS/NIBSC specification (>2.2 times 1011 platelets per pack) was achieved by 99% of PCS Plus and 82% of Auto C procedures. The positive correlation found between donor precount and final platelet yield was better for the PCS Plus. Both machines met U.K. specification for white-cell contamination but this was significantly greater for the Auto C. Plasma yields were similar.
As a result of this study we chose to use the PCS Plus for routine plateletpheresis in our unit. This has enabled us not only to comply with UK BTS/NIBSC specifications for apheresis platelets easily and cost effectively but also to meet our own higher specification (2.75 times 1011 platelets per pack) using existing staff and without extending the working day.
Keywords:cost effectiveness  donor platelet count  platelet yield  plateletpheresis  quality assurance  specification